Remote Commands
R&S
®
FSW
803User Manual 1173.9411.02 ─ 19
//If the results are not accurate, change the position of the
//the temporary markers manually.
//----Changing the position of the temp markers-----
CALC:MARK:X 100MHZ
//Positions the reference marker on 100 MHz.
CALC:DELT2:X 5KHZ
//Positions delta marker 2 and 3 at a distance of 5 kHz to the reference marker.
CALC:DELT3:X 1KHZ
//Corrects the position of delta marker 3 by 1 kHz.
CALC:MARK:FUNC:MDEP:RES?
//Queries the measurement results for the repositioned markers.
11.5.13 Remote Commands for EMI Measurements
The following commands are required to perform EMI measurements in a remote envi-
ronment. This measurement requires the R&S FSW-K54 option.
The following tasks specific to the EMI application are described here:
● Activating EMI Measurement................................................................................ 803
● Configuring EMI Markers...................................................................................... 803
● Configuring the EMI Final Test..............................................................................805
● Configuring EMI Limit Lines.................................................................................. 806
● Controlling LISN....................................................................................................806
● Retrieving EMI Results..........................................................................................807
● Evaluating the Results.......................................................................................... 809
● Programming Example: EMI Measurement.......................................................... 809
11.5.13.1 Activating EMI Measurement
EMI measurement must be activated explicitely.
CALCulate:MARKer:FUNCtion:FMEasurement:STATe...................................................... 803
CALCulate:MARKer:FUNCtion:FMEasurement:STATe <State>
This command turns the EMI measurement marker functionality on and off.
Parameters:
<State> ON | OFF
Manual operation: See "EMI" on page 115
11.5.13.2 Configuring EMI Markers
The commands required to configure EMI markers are described here.
Configuring and Performing Measurements