Remote Commands
R&S
®
FSW
847User Manual 1173.9411.02 ─ 19
INPut:EATT <Attenuation>
This command defines an electronic attenuation manually. Automatic mode must be
switched off (INP:EATT:AUTO OFF, see INPut:EATT:AUTO on page 847).
If the current reference level is not compatible with an attenuation that has been set
manually, the command also adjusts the reference level.
Parameters:
<Attenuation> attenuation in dB
Range: see data sheet
Increment: 1 dB
*RST: 0 dB (OFF)
Example:
INP:EATT:AUTO OFF
INP:EATT 10 dB
Manual operation: See "Using Electronic Attenuation (Option B25)" on page 398
INPut:EATT:AUTO <State>
This command turns automatic selection of the electronic attenuation on and off.
If on, electronic attenuation reduces the mechanical attenuation whenever possible.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
INP:EATT:AUTO OFF
Manual operation: See "Using Electronic Attenuation (Option B25)" on page 398
INPut:EATT:STATe <State>
This command turns the electronic attenuator on and off.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
INP:EATT:STAT ON
Switches the electronic attenuator into the signal path.
Manual operation: See "Using Electronic Attenuation (Option B25)" on page 398
11.7.3.3 Configuring a Preamplifier
INPut:GAIN:STATe........................................................................................................ 848
INPut:GAIN[:VALue].......................................................................................................848