Manual operation: See "Delete Line" on page 531
CALCulate:LIMit<k>:STATe <State>
This command turns the limit check for a specific limit line on and off.
To query the limit check result, use CALCulate:LIMit<k>:FAIL?.
Note that a new command exists to activate the limit check and define the trace to be
checked in one step (see CALCulate:LIMit<k>:TRACe<t>:CHECk on page 986).
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:LIM:STAT ON
Switches on the limit check for limit line 1.
Usage: SCPI confirmed
Manual operation: See "Disable All Lines" on page 531
CALCulate:LIMit<k>:TRACe<t>:CHECk <State>
This command turns the limit check for a specific trace on and off.
To query the limit check result, use CALCulate:LIMit<k>:FAIL?.
Note that this command replaces the two commands from previous signal and spec-
trum analyzers (which are still supported, however):
● CALCulate:LIMit<k>:TRACe on page 1085
● CALCulate:LIMit<k>:STATe on page 986
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:LIM3:TRAC2:CHEC ON
Switches on the limit check for limit line 3 on trace 2.
Manual operation: See "Traces to be Checked" on page 530
Checking the Results of a Limit Check
CALCulate:LIMit:CLEar[:IMMediate].................................................................................986
CALCulate:LIMit<k>:FAIL?..............................................................................................987
CALCulate:LIMit:CLEar[:IMMediate]
This command deletes the result of the current limit check.
The command works on all limit lines in all measurement windows at the same time.