Remote Commands
R&S
®
FSW
987User Manual 1173.9411.02 ─ 19
Example:
CALC:LIM:CLE
Deletes the result of the limit check.
Usage: SCPI confirmed
CALCulate:LIMit<k>:FAIL?
This command queries the result of a limit check.
Note that for SEM measurements, the limit line suffix <k> is irrelevant, as only one spe-
cific SEM limit line is checked for the currently relevant power class.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 694.
Return values:
<Result> 0
PASS
1
FAIL
Example:
INIT;*WAI
Starts a new sweep and waits for its end.
CALC:LIM3:FAIL?
Queries the result of the check for limit line 3.
Usage: Query only
SCPI confirmed
Manual operation: See "Spectrum Emission Mask" on page 113
See "Limit Check <n>" on page 213
See "Limit Check" on page 246
Programming Example: Using Limit Lines
The following examples demonstrate how to work with limit lines in a remote environ-
ment.
Example: Configuring Limit Lines..........................................................................987
Example: Performing a Limit Check......................................................................989
Example: Configuring Limit Lines
This example demonstrates how to configure 2 limit lines - an upper and a lower limit -
for a measurement in a remote environment.
//------------- Configuing the limit lines ---------------------
CALC:LIM1:NAME 'FM1'
//Names limit line 1 'FM1'.
CALC:LIM1:CONT:MODE ABS
//Selects absolute scaling for the horizontal axis.
CALC:LIM1:CONT 1 MHz,50MHz,100 MHz,150MHz,200MHz