Remote Commands
R&S
®
FSW
858User Manual 1173.9411.02 ─ 19
Manual operation: See "Trigger Source" on page 427
See "External Trigger 1/2/3" on page 428
See "Video" on page 428
See "IF Power" on page 429
See "RF Power" on page 429
See "Power Sensor" on page 430
[SENSe:]SWEep:EGATe:TYPE <Type>
This command selects the way gated measurements are triggered.
Parameters:
<Type> LEVel
The trigger event for the gate to open is a particular power level.
After the gate signal has been detected, the gate remains open
until the signal disappears.
Note: If you perform gated measurements in combination with
the IF Power trigger, the R&S FSW ignores the holding time for
frequency sweep, FFT sweep, zero span and I/Q mode meas-
urements.
EDGE
The trigger event for the gate to open is the detection of the sig-
nal edge.
After the gate signal has been detected, the gate remains open
until the gate delay is over.
*RST:
EDGE
Example:
SWE:EGAT:TYPE EDGE
Manual operation: See "Gate Mode" on page 433
11.7.4.3 Configuring the Trigger Output
The following commands are required to send the trigger signal to one of the variable
TRIGGER INPUT/OUTPUT connectors. The tasks for manual operation are described
in "Trigger 2/3" on page 383.
OUTPut:TRIGger<port>:DIRection...................................................................................858
OUTPut:TRIGger<port>:LEVel........................................................................................ 859
OUTPut:TRIGger<port>:OTYPe...................................................................................... 859
OUTPut:TRIGger<port>:PULSe:IMMediate.......................................................................860
OUTPut:TRIGger<port>:PULSe:LENGth.......................................................................... 860
OUTPut:TRIGger<port>:DIRection <Direction>
This command selects the trigger direction.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Setting Basic Measurement Parameters