Remote Commands
R&S
®
FSW
808User Manual 1173.9411.02 ─ 19
CALCulate:MARKer<m>:FUNCtion:FMEasurement:RESult? <Result>
This command queries the result of the EMI measurement at the marker position.
Return values:
<Result> Power level. The unit depends on the one you have currently
set.
Example:
CALC:MARK1:FUNC:FME:RES?
Queries the result of marker 1.
Usage: Query only
CALCulate:
DELTamarker<m>:FUNCtion:FMEasurement:LIMit<k>:CONDition ?
<Condition>
This command queries the condition of a deltamarker position in relation to a certain
limit line.
Return values:
<Condition> 0
The deltamarker has passed the limit check.
1
The deltamarker is inside the margins of a limit line.
2
The deltamarker has failed the limit check.
Example:
CALC:DELT2:FUNC:FME:LIM2:COND?
Queries the condition of deltamarker 2 in relation to limit line 2.
Usage: Query only
CALCulate:MARKer<m>:FUNCtion:FMEasurement:LIMit<k>:CONDition ?
<Condition>
This command queries the condition of a marker position in relation to a certain limit
line.
Return values:
<Condition> 0
The marker has passed the limit check.
1
The marker is inside the margins of a limit line.
2
The marker has failed the limit check.
Example:
CALC:MARK1:FUNC:FME:LIM2:COND?
Queries the condition of marker 1 in relation to limit line 2.
Usage: Query only