Remote Commands
R&S
®
FSW
876User Manual 1173.9411.02 ─ 19
This command is only available with option B21 (External Mixer) installed.
Usage: Query only
[SENSe:]CORRection:CVL:CLEAr
This command deletes the selected conversion loss table. Before this command can
be performed, the conversion loss table must be selected (see [SENSe:
]CORRection:CVL:SELect on page 878).
This command is only available with option B21 (External Mixer) installed.
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:CLE
Usage: Event
Manual operation: See "Delete Table" on page 373
[SENSe:]CORRection:CVL:COMMent <Text>
This command defines a comment for the conversion loss table. Before this command
can be performed, the conversion loss table must be selected (see [SENSe:
]CORRection:CVL:SELect on page 878).
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Text>
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:COMM 'Conversion loss table for
FS_Z60'
Manual operation: See "Comment" on page 375
[SENSe:]CORRection:CVL:DATA <Freq>,<Level>
This command defines the reference values of the selected conversion loss tables.
The values are entered as a set of frequency/level pairs. A maximum of 50 frequency/
level pairs may be entered. Before this command can be performed, the conversion
loss table must be selected (see [SENSe:]CORRection:CVL:SELect
on page 878).
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Freq> numeric value
The frequencies have to be sent in ascending order.
<Level>