n Checking the memoryMEMORY CHECK
Runs the memory check on ROM and RAM.
Pressing [Enter] starts the memory check.
After the check, the result is displayed.
n Checking the leak sensorLEAK SENSOR CHECK
Performs the operation test for the leak sensor.
1
Use a syringe filled with water to wet
the sensor unit at the bottom of the
leak sensor.
2
Wait about 10 seconds and then press
[Enter].
If the sensor detects a leakage, [Good] will
be shown.
If not, [No Good] will be shown.
3
Press [CE].
The result display will be cleared.
If the result is [No Good], adjust the detection level with [LEAK THOLD] function in the
Calibration Support Group.
"Setting the operation level of leak sensor
LEAK THOLD" P.66
After wetting and testing the leak sensor, wipe away the water on the tray
completely. When wiping away the water, do not apply pressure to the leak
sensor, the sensitivity decreases if the sensor unit touches the wall.
"4.14 Cleaning of the Leak Tray" P.122
54
LC-40D X3 CL LC-40D XR CL
3 Operation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu LC-40D XR CL and is the answer not in the manual?

Shimadzu LC-40D XR CL Specifications

General IconGeneral
BrandShimadzu
ModelLC-40D XR CL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals