Setting the subnet mask《SUBNET MASK》
Sets subnet mask of the controller which
controls the instrument.
Enter the value using the numeric keypad
and press [Enter].
Setting the default gateway《DEFAULT GATEWAY》
Sets the default gateway of the controller
which controls the instrument.
Enter the value using the numeric keypad
and press [Enter].
Setting the transmission mode《TRS MODE》
Select the communication distribution
when the controller is connected to an LC
workstation.
Enter the value using the numeric keypad
and press [Enter].
Set Value Description
0 Connects in manual setting at the system controller.
2 Can not be used.
3 Connects to LabSolutions.
Do not set values other than the set value above. Otherwise the instrument will
not work properly.
n
Setting the operation level of leak sensor《LEAK THOLD》
Enables to set the threshold value for the
leak sensor.
When an error occurs in the leak sensor
check, set the value again.
Calibration of the leak sensor must be performed by the system administrator.
The leak sensor detects liquid drops when the mobile phase leaks from the
plunger seal or tubes. When the sensor output value exceeds the predetermined
"threshold," the sensor issues an error and stops pumping.
1
Dip the sensor unit at the bottom of the leak sensor into water.
Upon detection of water, the leak sensor value (Actual Level) in the screen increases.
2
Read the maximum value of the leak sensor value (Actual Level).
Example) The maximum value here is "160".
66
LC-40D X3 CL LC-40D XR CL
3 Operation