3
Wipe off the water near the leak sensor.
The leak sensor value (Actual Value) will decrease.
When wiping away the water, do not apply pressure to the sensor unit of the
leak sensor. The sensitivity decreases if the sensor unit touches the instrument
wall surface.
4
Calculate the threshold of the leak sensor based on the value read in step 2.
Formula to calculate the leak sensor threshold:
Threshold = (Max. value-100)×0.7+100
Example) The value obtained here is "142".
5
Press [Enter], input the value obtained
in step 4, and press [Enter].
The new "threshold" is stored.
n Setting the flow rate correction parameter (ALPHA)《FLOW COMP FACTOR》
Not used in CL model.
Recalibration operation is executed by service personnel.
This allows the service personnel to set the
flow rate correction parameter.
n Setting the pressure sensor sensitivity correction factor《PRESS COMP FACTOR》
Not used in CL model.
Do not change the factory default value.
This allows service personnel to set the
pressure sensor sensitivity correction factor.
LC-40D X3 CL LC-40D XR CL
67
3.6 VP Functions Screen
3