Set the acceptance criterion of the
flow rate accuracy in "ML".
The value can be set between 0.001 mL
and 1.000 mL.
For example, if the set flow rate is 1 mL/min and the criterion is ±2 %, input 0.02 mL.
7
Input the measurement time (seconds).
The flow rate accuracy (%) and pass/fail
result are displayed.
To use the gravimetric method
1
Measure the weight (g) before delivery.
2
Set the flow rate and start delivery.
3
After confirming the delivery stabilizes, accumulate water in a measuring flask
for a specific time.
4
Measure the weight (g) before delivery.
5
In the [FLOW CHECK] screen, press
[Enter].
6
In the measurement method selection
screen, input "1".
7
Input the measured flow rate.
The value can be set between 0.1 mL/min
and 3.0 mL/min.
LC-40D X3 CL LC-40D XR CL
57
3.6 VP Functions Screen
3