To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
Set the acceptance criterion of the
flow rate accuracy in mL.
The value can be set between 0.001 mL
and 1.000 mL.
For example, if the set flow rate is 1 mL/min and the criterion is ±2 %, input 0.02 mL.
9
Input the weight (g) measured before
delivery.
10
Input the measurement time (seconds).
11
Input the weight (g) measured after
delivery.
The flow rate accuracy (%) and pass/fail
result are displayed.
To use a flow meter
1
Set the flow rate and start delivery.
2
After confirming the delivery stabilizes, measure the flow rate.
3
In the [FLOW CHECK] screen, press
[Enter].
4
In the measurement method selection
screen, input "2".
58
LC-40D X3 CL LC-40D XR CL
3 Operation

Table of Contents