5
Input the measured flow rate.
The value can be set between 0.1 mL/min
and 3.0 mL/min.
6
Set the acceptance criterion of the
flow rate accuracy in mL.
The value can be set between 0.001 mL
and 1.000 mL.
For example, if the set flow rate is 1 mL/min and the criterion is ±2 %, input 0.02 mL.
7
Input the flow rate measured with a
flow meter.
The flow rate accuracy (%) and pass/fail
result are displayed.