n Checking operation of the maximum and minimum pressure functions 《PRESS LIMIT
CHECK》
When checking operation, connect a resistance tube to the pump outlet to create 5 MPa
to 10 MPa pressure at 1 mL/min during delivery.
1
Press [Enter].
Start delivery at 1 mL/min. The pressure is
monitored for 1 minute and the maximum
pressure (P.MAX) and minimum pressure
(P.MIN) are automatically set.
Then the flow rate is automatically set to
1.5 mL/min and 0.5 mL/min, and whether
the P.MAX and P.MIN functions work
correctly is checked.
The result is displayed after checking is
completed.
60
LC-40D X3 CL LC-40D XR CL
3 Operation