Chapter 2 Outline
Mobile DaRt Operation Manual
2-4
2.3 Use Conditions
2.3.1 Environmental Conditions
Use environment
Ambient temperature : 10°C ~ 35°C
Relative humidity : 30% ~ 75% (no dew condensation)
Atmospheric pressure : 700 hPa ~ 1060 hPa
Storage environment (not for transport and storage before delivery)
Ambient temperature : -10°C ~ 40°C
Relative humidity : 30% ~ 60% (no dew condensation)
Atmospheric pressure : 700 hPa ~ 1060 hPa
CAUTION
Charge the battery in the use environment.
Do not charge in the storage environment.
Power supply
AC power
System : Single phase AC
Frequency : 50/60 Hz
Standard voltages : 100, 110, 120, 200, 220, 230, 240 V
(set one of the voltages when installing)
Voltage variation range : ±10% of standard voltages
Supply capacity : 1kVA
Power Supply Impedance : 100, 110, 120 V : No more than 1.0Ω
200, 220, 230, 240 V : No more than 4.0Ω
Earth
Earth terminal : Earth resistance of no more than 100Ω
Additional earth terminal : Earth resistance of no more than 100Ω