2.5 Classification of Equipment
Mobile DaRt Operation Manual
2-7
2.5 Classification of Equipment
The method of protection against electric shock
Internally powered equipment (when powered by the battery)
Class I equipment (during charging)
The degree of protection against electric shock
Type B equipment
The degree of protection against harmful ingress of water
Ordinary equipment
CAUTION
The main body of this system is not protected against immersion
into a liquid. Never use this unit in a place in which immersion into
a liquid may occur. Never spill liquid on the surface or in the inside
of this system. Otherwise, electrical shock may occur. When liquid
is spilt, contact your nearest authorized Shimadzu service center.
The degree of safety of application in the presence of a flammable
Anesthetic mixtures with air or with oxygen or nitrous oxide
Equipment is not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide
WARNING
Risk of explosion if used in the presence of flammable anesthetics.
The mode of operation
Continuous operation with intermittent loading