4.6 How to Change Exposure Condition
Mobile DaRt Operation Manual
4-23
An example of the selection
The case where the examinee has a standard body and the exposure is made
on the front (AP) of the Abdomen.
1. Selecting the filming region: Press [Abdomen] switch
2. Selecting the filming orientation: Press [Front] switch
3. Selecting the body thickness correction: Press [Normal] switch
NOTE
For detailed DR system imaging region registration, refer to Canon’s CXDI Manual.
For anatomical program numbers that correspond to the DR system imaging
region, refer to “9.7 Anatomical Program Number”
4.6.4 Registration of Anatomical Programs
Stored contents
The following factors for filming are stored in an anatomical program:
(Refer to “Table 3-1 Description and basic function of each switch”.)
Filming region
Filming orientation
Body thickness correction
Exposure condition X-ray tube voltage
Current time product
NOTE
The longest X-ray exposure time of the DR radiography is 1000msec.
Registed exposure conditions in the anatomical program used to X-ray
exposure by the use of the DR radiography so that the exposure time may
become 1000msec or less.

Table of Contents

Other manuals for Shimadzu MUX-100D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu MUX-100D and is the answer not in the manual?

Shimadzu MUX-100D Specifications

General IconGeneral
BrandShimadzu
ModelMUX-100D
CategoryMedical Equipment
LanguageEnglish

Related product manuals