Mobile DaRt Operation Manual
7
WARNING
CONSIDERATION FOR MECHANICAL SAFETY
Be careful when operating this equipment. Be aware operators and patient’s body
- hand, foot, etc. - may not be got jammed between moving parts of the system or
between the unit and other equipment when moving a part of the unit.
WARNING
NEVER USE THIS SYSTEM AT A PLACE WHERE FLAMMABLE OR
EXPLOSIVE GASES MAY BE PRESENT.
This system is not explosion-proof. Never use this equipment at a place where
flammable or explosive gases may be present.
WARNING
ABOUT ELECTRICAL EARTHING
Connect the power cord plug to a socket with an electrical earthing terminal.
CAUTION
DO NOT SPLASH WATER ON THIS EQUIPMENT !
Do not splash water on this equipment, because there is a risk of electric shock.
To clean the equipment, wipe the surfaces with a cloth dampened with
antibacterial solution.
CAUTION
THE SYSTEM MAY FALL OVER, IF IT IS OPERATED ON A SURFACE
INCLINED MORE THAN 5 DEGREES.
When operating the tube arm of the equipment make sure the equipment is
located on level ground or floor where the slope is less than 5 degrees.
CAUTION
The operator should set the focal spot to skin distance as large as practible to
ensure the absorbed dose to the patient is as low as possible.
CAUTION
PAY ATTENTION TO OBSTACLES DURING DRIVING THE UNIT!
If the unit hits obstacles during driving and the force applied to the bumper switch
by the hitting is less than 2 kg, the front bumper switch will not detect the obstacle.