Chapter 4 How to Operate
Mobile DaRt Operation Manual
4-32
Fig. 4-16 First position of the hand switch
NOTE
The light is not illuminated unless the key is turned off or the tube arm is
latched.
Only if the stand by illuminates, exposure preparation can be made. Confirm
that the stand by illuminates before pushing the hand switch.
NOTE
During preparation, the collimator lamp will turn off automatically.
When preparation has been performed, the collimator lamp will not illuminate
until the exposure is complete. If the collimator lamp needs to be turned on,
the preparation/exposure button must be released. When the collimator lamp
is to be lighted after Ready-up, refer to “4.9 User Setting”.
3. Preparation will be completed and Ready indicator lamp will come on.
Operational state indicator lamp is Green.
NOTE
If the X-ray exposure is not made within 30 seconds after the handswitch
has been pressed to the first position, the preparation will be automatically
cancelled.
To prepare an X-ray exposure again, repeat procedure above.
If the Collimator lamp lights after Ready up, refer to “4.9 User Setting”.

Table of Contents

Other manuals for Shimadzu MUX-100D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu MUX-100D and is the answer not in the manual?

Shimadzu MUX-100D Specifications

General IconGeneral
BrandShimadzu
ModelMUX-100D
CategoryMedical Equipment
LanguageEnglish

Related product manuals