Commissioning
FLOWSIC500 · Operating Instructions · 8025733/1GMJ/V4-2/2022-07 · © SICK Engineering GmbH 77
Subject to change without notice
4.3.4 Configure the power management
Select the type of power supply corresponding to the configuration of the FLOWSIC500:
Dynamic Mode (External + Backup):
Measuring rate: 4 Hz
Battery Mode (2 x Battery):
Measuring rate: 1 Hz, to maximize the battery service life
Eco Mode (External + Backup):
Standard setting: Measuring rate is 4 Hz when the external voltage supply is available.
The measuring rate is set automatically to 1 Hz if the external voltage supply fails in
order to maximize the backup battery service life.
Fig. 44 Power management