82 FLOWSIC500 · Operating Instructions · 8025733/1GMJ/V4-2/2022-07 · © SICK Engineering GmbH
Operation
Subject to change without notice
5.2.3 Main screen (without device option volume conversion)
Use buttons and to toggle between the menu entries on a level.
Press ENTER to go down one menu level.
Main display
The following information is displayed on the top display menu level:
Main display Description
V
000000000 m
3
20.08.2021 10:30:52
V
= Volume absolute, cannot be reset
Pressing ENTER opens the FLOWSIC500 menu.
err
V
000000000 m
3
20.08.2021 10:30:52
err
V = Error volume: Volume counted during a malfunction, can be
reset
Pressing ENTER opens the action “Reset error volume”.
“Resetting the error volume” (p. 98).
Q 0.000 m3/h
VOG 0.000 m/s
Q = Volume flow
VOG = Gas velocity
Current events
1 Event
Current events (1 event is reported)
Pressing ENTER opens a list of current reported events.
Use the arrow buttons to toggle through the reported events.
Event Summary
2 Events
Stored status messages: Events since the last time the event
summary was reset (2 events have occurred).
Pressing ENTER opens a list of stored events.
Use the arrow buttons to toggle through the stored events.
Last Event Reset
20.08.2021 10:30:52
Last reset of event summary
Pressing ENTER opens the action “Reset Event Summary”.
“Resetting the event summary” (p. 98).
NOTICE:
A parameter with Malfunction status is shown on the display by a blinking
exclamation mark after the parameter (e. g. Q!).