Installation
FLOWSIC500 · Operating Instructions · 8025733/1GMJ/V4-2/2022-07 · © SICK Engineering GmbH 65
Subject to change without notice
3.5.3 Installing the temperature sensor
1 Insert the temperature sensor into the immersion sleeve to the stop.
2 Tighten the locknut.
3 Have the wire seal attached by a calibration inspector (→ Fig. 10).
Fig. 35 Installing the temperature sensor
3.6
Fitting a display protector (option)
Display protector (Part No. 2085547) is available as an option for protection of the display
from UV light.
Fig. 36 Display protection
NOTICE:
SICK recommends to install the temperature sensor on the temperature
measuring port which is on the same side as the display.
The temperature sensor can be greased with heat-conductive oil or paste to
improve its performance.
1 Temperature sensor
1