8024638/AE00/V1-0/2019-09| SICKSERVICE MANUAL | MCS200HW
Subject to change without notice
5 REPAIRS
Fig. 224: Firmware update 2
5.4.3.2 Swapping the SD card
If a parameter backup could not be created, proceed as follows:
1 Disconnect the MCS200HW power supply.
2 Insert the SD card of the old LINUX CPU into the new LINUX CPU.
3 Connect the MCS200HW power supply.
4 Open the SOPAS menu.
5 Select Maintenance -> Load/Save parameter.
6 Select latest version (with latest timestamp).
7 Click Restore parameter backup to confirm (see Fig. 224 “Firmware update 2”,
page 138).