48
8024638/AE00/V1-0/2019-09| SICKSERVICE MANUAL | MCS200HW
Subject to change without notice
4 FUNCTIONAL TEST
1 Select Parameterization -> Measuring components -> Definition measuring
components.
2 Compare individual warning thresholds of the individual components against control
values of the reference energy. The reference energy of the respective component
must be at least 5 percentage points above the individual warning threshold.
3 Proceed as follows when the reference energy is lower or a maintenance request is
pending:
1 Check the beamer current (see Chapter “Beamer current check”, page 47).
2 When the beamer current is OK, check optical components:
Window soiling
Soiling of the gold layer of the mirror plates
–Cloudiness of the lens
4.5 Test signal transmission
4.5.1 Setting the IP address
See SOPAS ET Manual.