82
8024638/AE00/V1-0/2019-09| SICKSERVICE MANUAL | MCS200HW
Subject to change without notice
5 REPAIRS
Fig. 116: Bundling the cables of the heating cartridges
10 Fasten the cables of the new heating cartridges with clamping screws in the contact plug
(
see Fig. 115 “Contact plug, assignment with circuit diagram”, page 81).
▸ Shorten the cables to the appropriate length.
11 Fasten the fastening clip in the contact plug.
12 Close the contact plug.
13 Fasten the cell covers (see Chapter “Fitting the complete cell”, page 68).
14 Fit the cell (see Chapter “Fitting the complete cell”, page 18).
5.2.6 Replacing the PT100 temperature sensor
To replace the PT100, the complete cell must be removed and all covers removed.
Heating cartridge pairs have the clamps next to each other.
Tools required
Phillips screwdriver
Slot screwdriver
Pliers
Spare parts set No. Contents
2099813 Measuring resistor PT100
DANGER: Risk of burns due to hot surface.
Allow the MCS200HW to cool down before carrying out any work.