8024638/AE00/V1-0/2019-09| SICK S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
FUNCTIONAL TEST 4
4.3 Beamer current check
1 Select Diagnosis -> State control -> Beamer.
2 Check beamer current. The beamer current must be within the specified warning
thresholds.
3 Replace the beamer when the radiator current is less than 0.05 A above the Current
Min. Warning or a maintenance request is pending.
4.4 Reference energy check
1 Select Diagnosis -> Control values -> Reference energy.
2 Check the reference energy.