8024638/AE00/V1-0/2019-09| SICK S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS 5
5.2.5 Replacing the heating cartridges
To replace the heating cartridges, the complete cell must be removed and all covers
removed.
Fig. 110: Heating cartridges
1 Remove the complete cell (see Chapter “Removing the complete cell”, page 65).
2 Place the cell down with the rear side upwards.
>>>
The side with the cell inlet filter and ejector is at the bottom.
Fig. 111: Cell, overview, left
3 Loosen the fastening clips of the heating cartridges with a Phillips screwdriver (9 pieces).
Tools required
Phillips screwdriver
Slot screwdriver
Pliers
Spare parts set No. Contents
2099812 Heating cartridge garlands set, 2-HP 115 V, 180 W, 10 x 80
DANGER: Risk of burns due to hot surface.
Allow the MCS200HW to cool down before carrying out any work.
1Heating cartridges
2Grounding
3 Cable to thermostatic switch
4 PT100
Document the arrangement of the cables of the heating cartridges for reassembly, e.g.
take a photo.

Table of Contents

Other manuals for SICK MCS200HW

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK MCS200HW and is the answer not in the manual?

SICK MCS200HW Specifications

General IconGeneral
BrandSICK
ModelMCS200HW
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals