Scale parameters and functions of the belt scale
8.9 DR 7 Process interfaces
SIWAREX WP241
104 Manual, 04/2014, A5E33713528A
By parameterizing the corresponding digital output you can use one of the digital outputs as
a pulse sensor/quantity (see Assignment digital output DQ 0, 1, 2, 3 (Page 100)).
In addition you must specify a load per pulse and a pulse duration for the signal. If the
material flow rate is displayed e.g. in t/h, you must specify the quantity per pulse in t, or with
kg/h in kg. The pulse duration is defined in ms.
Make sure when setting the parameters that you only enter plausible combinations → pulse
duration > (amount per pulse / material flow rate)