Messages
9.5 Message list
SIWAREX WP241
Manual, 04/2014, A5E33713528A
141
The message list is an overview of all messages that the SIWAREX module can generate. A
message can be quickly identified by the message code (number).
System status message list
Operating errors (code 1000 to 1999)
sorted by code no.
1000 operating error exists
Group message, at least one operating error exists.
1001 Watchdog 1001 Watchdog, error is displayed for at least 10 seconds. A
serious error has occurred in the function of SIWAREX, e.g.
program error, severe electromagnetic influence on device,
etc.
Contact the SIWAREX Support if the error occurs multiple
1002 RAM error 1002 RA
M error. An error has occurred in the memory; the memory
content is no longer correct. The module must be switched
off. If the error occurs again, SIWAREX is defective.
1003 Checksum incorrect parameter 1003 Checksum error at parameter. Critical error because the
parameters are no longer safe.
1004 Checksum incorrect program 1004 Checksum error program code. Critical error because the
program is no longer safe.
Error when writing/deleting,or logbook full
1102 ADU error 1102 AD converter error when reading in the measured value. If
the error occurs again, make sure that the EMC
recommendations are observed (chapter EMC-compliant
setup (Page 21)).
Undervoltage at sensor cables
Overload of scale (ca. 110%)
Underload of scale (ca. -10%)
1107 Legal trade display failure 1107 The SecureDisplay legal trade display no longer
communicates with the module