Connecting
6.3 Connecting the load cells
SIWAREX WP241
Manual, 04/2014, A5E33713528A
27
Rules
Observe the following rules when connecting analog (strain gauge) load cells:
The load cells fitted in the belt scale are connected in parallel in a junction box (e.g.
SIWAREX JB). Use a well-shielded cable for the connection between junction box and
SIWAREX WP241 (see SIWAREX cables Accessories (Page 177)).
1. The cable shield is always applied at the cable gland of the junction box (SIWAREX JB).
If there is a risk of equipotential bonding through the cable shield, connect an
equipotential bonding conductor parallel to the load cell cable.
2. Twisted wire pairs that are also shielded are required for the specified cables:
Sensor cable (+) and (-)
Measuring voltage cable (+) and (-)
Supply voltage cable (+) and (-)
Figure 6-2 Shielding in the screw gland
We recommended that you use the cables listed in chapter → Accessories (Page 177).
3. The shield must be connected to ground directly in the vicinity of the SIWAREX WP241.
The maximum distance between the SIWAREX WP241 and the load cell applies when
using the recommended cables (see SIWAREX cablesAccessories (Page 177)).
Table 6- 2 Load cell connections on the module
Labeling
Function
Sig-
Measurement cable load cell -
Sig+
Measurement cable load cell -
Sen-
Sensor cable load cell -
Sen+
Sensor cable load cell +
Exc-
Supply load cell -
Exc+
Supply load cell +

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIWAREX WP241 and is the answer not in the manual?

Siemens SIWAREX WP241 Specifications

General IconGeneral
BrandSiemens
ModelSIWAREX WP241
CategoryAccessories
LanguageEnglish

Related product manuals