11.2 Restricting the maximum mattress base height
11.2.1 On the LCD handset
The mattress base must be horizontal before a restriction can be defined.
A pre-set maximum mattress base height cannot be exceeded in the daily care routine or by
the patient. When the bed is raised, it only moves to this position and then stops. A maximum
mattress base height can only be defined by service personnel or qualified medical staff who
have been trained by the operator to do so.
1 Set the bed to the position that you wish to define as the maximum mattress base height.
2 Press the UP and DOWN buttons si-
multaneously, and keep them pressed
while you hold the top end of the
handset briefly (for about ½ second)
against the orange magnetic unlock-
ing key.
3 Keep both the UP and DOWN buttons
pressed for a further 5 seconds while
the warning triangle symbol is dis-
played.