17.3 Inspection report
Table zz: Test specifications for the immunity of sheathings to high-frequency wire-
less communication equipment
Test Fre-
quency
MHz
Frequen-
cy band
Radio
service
Modula-
tion
Max.
power W
Distance
m
Immunity
test level
v/m
1720 1700 to
1990
GSM 1800
CDMA
1900, GSM
1900,
DECT, LTE
band 1;3; 4;
25; UMTS
Pulse mod-
ulation 217
Hz
2 0.3 28
1845
1970
2450 2400 to
2570
Bluetooth,
WLAN
802.11
b/g/n, RFID
2450, LTE
band 7
Pulse mod-
ulation 217
Hz
2 0.3 28
5240 5100 to
5800
WLAN
802.11 a/n
Pulse mod-
ulation
217Hz
0.2 0.3 9
5500
5785
Inspection of Electromedical Devices according to DIN EN 62353 (VDE 0751-1):
2015-10
Customer/med. facility/practice:
Address:
Carried out: [ ] Repeat inspection [ ] Inspection prior to initial op-
eration(reference value)
[ ] Inspection following repairs/maintenance
Equipment type: [X] Hospital
bed
[ ] Care bed Protection class: [X] I [ ] II
Bed type: Seta pro Inventory number:
Location: Serial number:
Appendix
149 Instruction Manual