12 Cleaning and disinfection
12.1 Safety information on cleaning and disinfection
12.1.1 Before starting cleaning
12.1.2 After cleaning
Failure to follow this safety advice could result in considerable damage to the bed and the
electrical components and lead to subsequent malfunctions!
Ensure that all plugs for the drive system are connected as prescribed.
Lock the electric adjustment functions.
Hang or insert all handsets in the holders provided on the bed.
Unplug the mains plug from the electrical socket.
Store the mains plug so that it does not come into excessive contact with cleaning solu-
tions.
Ensure that none of the electrical components show any signs of external damage.
If these instructions are disregarded, water or cleaning agents may penetrate the sys-
tem resulting in malfunctions or damage.
Before operating the bed again, ensure that there is no residual moisture on the electri-
cal contacts by drying or blowing on the power plug.
If you suspect that water or cleaning solutions have penetrated into electrical compo-
nents:
Unplug the bed immediately from the power socket and do not insert the power
plug into the socket again,
Label the bed clearly as “OUT OF ORDER”,
Take it out of service immediately and report this to the operator without delay.
Instruction Manual 108