17.2.1 EMC information tables
ATTENTION
Risk of malfunctions
Failure to heed this warning may result in malfunctions of the bed or of surrounding devices.
•
The bed is intended for use in the electromagnetic environment described below. The op-
erator or user of the bed should ensure that it is used in such an environment.
Note: The characteristics of this device, determined by emissions, allow its use in industry
and hospitals (CISPR 11, Class A). This equipment may not provide adequate protection for
radio services when used in a residential environment (which typically requires Class B ac-
cording to CISPR 11). If necessary, the user must take remedial measures such as imple-
menting or realigning the device.
Ambient limit values of the interference emissions
Phenomenon Professional healthcare facilities
Conducted and radiated interference emissions CISPR 11 Group 1, Class A
Sheathing
Phenomenon EMC basic standard or
test method
Immunity test level
Professional healthcare
facilities
Electrostatic discharge (ESD) IEC 61000-4-2 +/- 8 kV contact
+/- 2 kV, +/- 4 kV ¸ +/- 8 kV, +/-
15 kV air
High-frequency electromagnet-
ic fields
IEC 61000-4-3 3 V/m; 80 MHz up to 2.7 GHz;
80% AM at 1 kHz
High-frequency electromagnet-
ic fields in the immediate vicini-
ty of wireless communication
devices
IEC 61000-4-3 See separate table zz at the
end of this chapter.
Appendix