Sheathing
Phenomenon EMC basic standard or
test method
Immunity test level
Professional healthcare
facilities
Magnetic fields with rated pow-
er frequencies
IEC 61000-4-8 30 A/m; 50 Hz or 60 Hz
AC port for supply input
Phenomenon EMC basic standard Immunity test level
Professional healthcare
facilities
Electrical fast transients/bursts IEC 61000-4-4 +/- 2 kV; 100 kHz repetition fre-
quency
Electrical surges: conductor to
earth
IEC 61000-4-5 +/- 0,5 kV; +/- 1kV; +/- 2 kV;
Electrical surges: conductor to
conductor
IEC 61000-4-5 +/- 0,5 kV; +/- 1kV
Conducted interference in-
duced by high-frequency fields
IEC 61000-4-6 3 V; 0.15 MHz to 80 MHz; 6V
in ISM frequency bands be-
tween 0.15 MHz and 80 MHz
80% AM at 1 kHz
Voltage dips IEC 61000-4-11 0% U
T
; ½ cycle; at 0, 45, 90,
135, 180, 225, 270 and 315
degrees
0% U
T
; 1 cycle; and 70% U
T
;
25/30 cycles; single-phase at 0
degrees Celsius
Voltage interruptions IEC 61000-4-11 0% U
T
; 250/300 cycles
Appendix
147 Instruction Manual