11.3 Restricting minimum mattress base height
11.3.1 Setting the minimum mattress base height (on the LCD handset)
The mattress base must be horizontal before a restriction can be defined.
A pre-set minimum mattress base height cannot be exceeded in the daily care routine or by
the patient. When the bed is lowered, it only moves to this position and then stops. A mini-
mum mattress base height must only be defined by technical staff or qualified medical staff
who have been trained by the operator to do so.
1 Set the bed to the position that you wish to define as the minimum mattress base height.
2 Press the UP and DOWN buttons si-
multaneously, and keep them pressed
while you hold the top end of the
handset briefly (for about ½ second)
against the orange magnetic unlock-
ing key at the foot end of the bed.
3 Keep both the UP and DOWN buttons
pressed for a further 5 seconds while
the warning triangle symbol is dis-
played.