6.5.13 Lead-acid battery (optional)
Button Explanation
Battery fully charged
Battery approx. 2/3 charged
Battery approx. 1/3 charged - connect bed to power supply as soon as pos-
sible
Battery empty - connect bed to power supply as soon as possible
Bed connected to power supply:
Control box indication Indication on the display Explanation
No indication on display The battery is full, or a fully de-
pleted battery is preparing for
charging
Yellow LED on, battery charg-
ing
A conventional high-capacity lead-acid battery can be used for temporary emergency opera-
tion of the electrical drive system. This guarantees that all electric adjustments can be carried
out even during a power cut.
Emergency operation
When the bed is occupied by a patient of normal weight (approx. 80 kg), up to 8 full UP/
DOWN adjustments can be made if the battery is new and fully charged.
Under emergency conditions, if the battery capacity is depleted, a signal tone will sound dur-
ing adjustments.
Product description
Instruction Manual 40