EasyManua.ls Logo

Stiegelmeyer Seta pro - Lead-Acid Battery (Optional)

Stiegelmeyer Seta pro
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.5.13 Lead-acid battery (optional)
Button Explanation
Battery fully charged
Battery approx. 2/3 charged
Battery approx. 1/3 charged - connect bed to power supply as soon as pos-
sible
Battery empty - connect bed to power supply as soon as possible
Bed connected to power supply:
Control box indication Indication on the display Explanation
No indication on display The battery is full, or a fully de-
pleted battery is preparing for
charging
Yellow LED on, battery charg-
ing
A conventional high-capacity lead-acid battery can be used for temporary emergency opera-
tion of the electrical drive system. This guarantees that all electric adjustments can be carried
out even during a power cut.
Emergency operation
When the bed is occupied by a patient of normal weight (approx. 80 kg), up to 8 full UP/
DOWN adjustments can be made if the battery is new and fully charged.
Under emergency conditions, if the battery capacity is depleted, a signal tone will sound dur-
ing adjustments.
Product description
Instruction Manual 40

Table of Contents

Related product manuals