CAUTION
Risk of injury and material damage
Failure to heed this warning may result in injuries and damage to property.
•
If any safety-relevant damage or malfunction is suspected during testing in accordance
with DIN EN 62353 (VDE 0751-1), take the bed out of service at once until it has been
repaired or the damaged components have been replaced! A repeat inspection must be
carried out in accordance with the test sheet to determine whether the damage or mal-
function has been rectified. Only then have the requirements for continued operation been
met.
An inspection sheet template for testing in accordance with DIN EN 62353 (VDE 0751-1) is
given in Inspection report » 149.
Maintenance
115 Instruction Manual