Stiegelmeyer Seta pro - Routine Bed Operation and Adjustments; Moving and Securing the Bed

Stiegelmeyer Seta pro
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAUTION
Health risks and material damage
Failure to heed this warning may result in health risks and damage to property!
The maximum loading capacity at the front end of the patient lifting pole is 45 kg.
Only use the patient lifting pole with the head section and grab handle also inserted (for
maximum stability of the patient lifting pole adapter sleeve)
Always leave the patient lifting pole inserted when children are left unattended in the
same room as the bed. In this way, any sharp edges on the patient lifting pole adapter
sleeve that may develop during use are protected against contact
The patient lifting pole is not suitable for rehabilitation exercises.
Pay attention to door clearances when moving beds with patient lifting poles or infusion
stands attached.
1 Attaching: Insert the long, straight end of
the patient lifting pole from above into the
adapter sleeve.
To remove: Pull the patient lifting pole
straight up and out of the adapter sleeve.
2 Make sure that the metal pins on the
patient lifting pole slot into the adapter
sleeve recesses and engage.
In this way, the patient lifting pole is
prevented from twisting sideways.
Bed reprocessing/bed adaptation
Instruction Manual 56