EasyManua.ls Logo

Svantek SVAN 971 - Function #5 - Statistical Analysis Results Read - out

Svantek SVAN 971
220 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SVAN 971 User Manual Appendixes
107
Response:
#D,e; - command was executed
#D,e,?; - command cannot be executed
8) #D,m,<address>,<dirName>; function enables the user to create a subdirectory in the directory
defined by <address>:
Response:
#D,m; - command was executed
#D,m,?; - command cannot be executed
9) #D,f,<address>; function enables the user to delete directory and its contents (files and
subdirectories):
Response:
#D,f; - command was executed
#D,f,?; - command cannot be executed
A.7 FUNCTION #5 - STATISTICAL ANALYSIS RESULTS READ-OUT
#5 function enables one to read out the statistical analysis results.
#5 function format is defined as follows:
#5,p;
where:
p - the number of the profile (1, 2 or 3)
The device responds, sending the current classes of the statistics in the following format:
#5,p;<Status Byte> <LSB of the transmission counter> <MSB of the transmission counter>
<NofClasses><BottomClass><ClassWidth><Counter of the class> (...) <Counter of the class>
Status Byte gives the information about the current state of the instrument.
D7
D6
D5
D4
D3
D2
D1
D0
where:
D7 = 0 means "overload does not happen",
= 1 means "overload appeared",
D6 = 1 reserved,
D5 = 0 instantaneous current result (RUN State),
= 1 final result (STOP State),
D0 to D4 reserved bits.
Note: There is not any succeeding transmission in the case when the Status Byte is equal to
zero.
The transmission counter is a two-byte word denoting the number of the remaining bytes to be
transmitted. Its value is calculated from the formulae:
Transmission counter = 6+n * (4 * the number of the classes in the statistics)

Table of Contents

Related product manuals