SVAN 971 User Manual
45
Field description of the One Result mode
1. Result name:
for SLM, 1/1 Oct. and 1/3 Oct. results: OVL,
Lpeak, Lmax, Lmin, L, Leq, LE, Lden, LEPd,
Ltm3, Ltm5, Lnn
for Dosimeter results: OVL, Lpeak, Lmax,
Lmin, L, DOSE, D_8h, PrDOSE, LAV, Leq, LE,
SEL8, E, E_8h, LEPd, PSEL, Ltm3, Ltm5, Lnn,
PTC, PTP, ULT, TWA, PrTWA, Lc-a
for Run. LEQ results: OVL, Lpeak, Lmax, Lmin,
L, Leq, LE, Lden, LEPd, Ltm3, Ltm5, Lnn,
LR15, LR60
2. Value of the measured result
3. Profile number
4. Quasi analogue value indicator
5. Implemented weighting filter: Z, A, C or B
6. Detector time constant, when the detector is exponential: Imp., Fast, Slow or Lin when the detector
is linear
7. Units of measured value
8. Elapsed time
Elapsed time shows the current second of the measurement. The value presented there belongs to the
range [0, Integration Period]
Note: For some results, weighting filters and detector type are presented in the result name.
For example, result Lmax with A filter and Fast detector will be presented as LAFmax. For
such results, there is no indication in the filter and detector field.
6.1.2 Three profiles display mode
In the 3 Profil. mode any three measurement results,
selected in the Disp. Res list, may be presented for three
profiles. You may change the 3 Profil. view by pressing
the <Enter> key.
<ENT>
Field description of the 3 Profiles mode
1. Result for the first profile
2. Result for the second profile
3. Result for the third profile
4. Implemented weighting filter: A, C, Z or B and
detector time constant: I (Impulse), F (Fast), S
(Slow) when the detector is exponential or L when
the detector is linear
1
2
3
5
1
2
6
4
3
7
8
4