Dosimeter specific parameters can be set in accordance with the OSHA HC
(Occupational Safety and Health Administration - Hearing Conversation),
OSHA PEL (Occupational Safety and Health Administration – Permissible
Exposure Level) and ACGIH standards.
Criterion Level is a steady noise level permitted for a full eight-hour work shift:
60dB, 65dB, 70dB, 75dB, 80dB, 84dB, 85dB, 87dB, 90dB;
Threshold Level is a noise level limit below which the dosimeter does not
accumulate noise dose data: None, 60dB, 65dB, 70dB, 75dB, 80dB, 85dB,
90dB;
Exchange Rate is an amount by which the permitted sound level may increase if
the exposure time is halved: 2, 3, 4, 5, 6;
ULT Threshold Level (Upper Level Time) is a threshold level for calculation of
ULT results: 70dB ÷ 140dB;
PTC Threshold Level (Peak Threshold Counter) is a threshold level for
calculation of PTC results: 70dB ÷ 140dB.
The Exp. Time enables setting the desired value of the
workday exposure time that is used for calculation of the
LEPd results (see Appendix D).
The Alarm position is active only in the Dosimeter
function and enables programming the alarm thresholds
for three profiles (Thresh. P1 (2,3)).
Alarm is signalled by a diode flashing. To stop the diode
flashing, press any key.
Thresholds can be set for next measurement results of
the Dosimeter in ranges:
DOSE: 1÷200%;
D_8h: 1÷200%;
PTC: 1÷1000;
ULT: 1÷60s.
If the Off value is selected, the alarm for the
measurement result is switched off.