SVAN 971 User Manual – Appendixes
Delay in the start of measurement
Ynn - nn delay given in seconds (0 59) and
(60 3600) with step 60s
Synchronization the start of
measurement with RTC
y0 - switched off (OFF)
y1 - synchronization to 1 min.
y15 - synchronization to 15 min.
y30 - synchronization to 30 min.
y60 - synchronization to 1 hour.
State of the instrument (Stop, Start or
Pause)
S0 - STOP
S1 - START
S2 - PAUSE
Threshold level for ULT calculation
Xlnnn:p - nnn level in dB (70 140)
p: 1, 2, 3 - profile number
Logger Triggering mode (TriggerMode)
XT0 - switched off (OFF)
XT4 - LEVEL +
XT5 - LEVEL –
Logger Triggering level (TriggerLev)
XLnnn - nnn level in dB (24 136)
Logger Triggering - Number of records
taken into account before the fulfilment
of the triggering condition (TriggerPre)
XQnn - nn number of records saved in the
logger before the triggering condition;
nn (0 10)
Logger Triggering - Number of records
taken into account after the fulfilment of
the triggering condition (TriggerPost)
Xqnnn - nnn number of records saved in
the logger after the fulfilment of the triggering
condition; nn (0 200)
Threshold level for PTC calculation
XCnnn:p - nnn level in dB (70 140)
p: 1, 2, 3 - profile number
Logger File Splitting Mode
XA0 switched off (OFF)
XA-1 - The file is created for each measurement
cycle.
XA15 - The file is created every 15 min.
synchronized to RTC.
XA30 - The file is created every 30 min
synchronized to RTC.
XA60 - The file is created every 1 hour
synchronized to RTC.
XA1440 - The file is created on the specified
times.
Specified Time for Logger File Splitting
XDnnn:p – nnn:
-1 - off.
0:1439 - Time in minutes.
Valid only if SplitMode is equal 1440.
p: 1 … 6 – specified time number