EasyManua.ls Logo

Tektronix TG8000

Tektronix TG8000
372 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ATG7 module adjustment procedures
15. Select the DAC G
ain (700 m V DC) calibration signal for outputs as follows:
a. Press the left () or right () arrow button to select BLACK 1,andthen
press the ENTE
R button.
b. Press the left ()orright() arrow button to select DAC Gain (700 mV
DC),andthe
npresstheENTER button.
c. Press the BACK button.
d. Repeat parts a through c of this step to select the calibration signal for
the BLACK 2, SIGNAL, and BARS outputs.
16. Locate variable resistor R481 on the ATG7 circuit board. (See Figure 5-1.)
17. Adjust R481 so that the output gain is 700 mV ±1 m V.
18. Change the BNC cable connection from the BLACK 1 connector to the
BLACK 2 connector on the ATG7 module.
19. Locate variable resistor R581 on the ATG7 circuit board. (See Figure 5-1.)
20. Adjust R581 so that the output gain is 700 mV ±1 m V.
21. Chang
e the BNC cable connection from the BLACK 2 connector to the
SIGNAL connector on the ATG7 module.
22. Locat
e variable resistor R 681 on the ATG7 circuit board. (See Figure 5-1.)
23. Adjust R681 so that the output gain is 700 mV ±1 m V.
24. Change the BNC cable connection from the SIGNAL connector to the BARS
connector on the ATG7 module.
25. Locate variable resistor R781 on the ATG7 circuit board. (See Figure 5-1.)
26. Adjust R781 so that the output gain is 700 mV ±1 m V.
TG8000 Multiformat Test Signal Generator Service Manual 5–5

Table of Contents

Other manuals for Tektronix TG8000

Related product manuals