DVG7 module adjustment procedures
10. Record the a tte
nuation factor value in the DVG7 output amplitude table.
11. Enter the attenuation factor value into the oscilloscope:
a. On the oscilloscope, press the SETUP D IALOG button.
b. Select the “Vert” tab from the top of the setup page.
c. Select the “External Attenuation” box, and then enter the attenuation
factorcalculatedinstep9above.
9–6 TG8000 Multiformat Test Signal Generator Service Manual