EasyManuals Logo
Home>Tektronix>Portable Generator>TG8000

Tektronix TG8000 Service Manual

Tektronix TG8000
372 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #229 background imageLoading...
Page #229 background image
DVG7 module adjustment procedures
DVG7 output amp
litude adjustment table. Print this table for use during the
adjustment procedure.
Table 9-2: DVG7 output amplitude
Minimum Value Maximum
Characterization
DMM Measure
ment
(typically 0.2880 V)
—— ——
Oscilloscope Cycle R MS
(typically 116 mV)
—— ——
Oscilloscope Cycle M ean
(typical
ly 1 mV)
—— ——
Sine wave
RMS amplitude
(typically 116 mV)
——
Attenua
tion Factor
2.35 2.55
ValueAdjustment Record
Minimum Before adjustment After adjustment Maximum
Signal 1 amplitude
Measu
red value (μ)
with attenuation
720 mV 880 mV
Black
1 amplitude
(Option BK only)
Measured value (μ)
with
attenuation
720 mV 880 mV
TG8000 Multiformat Test Signal Generator Service Manual 9–7

Table of Contents

Other manuals for Tektronix TG8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TG8000 and is the answer not in the manual?

Tektronix TG8000 Specifications

General IconGeneral
BrandTektronix
ModelTG8000
CategoryPortable Generator
LanguageEnglish

Related product manuals