Preventive Maintenance Guidelines
Thermo Scientific Smart-Vue
®
Wireless Monitoring Solution Administrator Guide v 2.2 129
22.1.4.3 Short-Press On single-sensor modules, a short-press on the button causes the module
to read the sensor refresh the display.
On dual-sensor modules, such as temperature/humidity and
temperature/CO
2
, the reading for sensor 1 is displayed for a few seconds
and then the reading for sensor 2 is displayed for a few seconds. The display
returns to its regular display of sensor 1.
22.1.4.4 Long-Press A long-press of three seconds or more on the Smart-Vue module button
initiates automated module installation into your monitoring network using
the integrated Service Discovery Protocol. The sensor is then visible within
the Smart-Vue Client application. The long-press also begins logging
readings in module memory if the module had been stopped. Datalogging
starts with the parameters already present in memory (in particular, the
reading interval, alarm limits, and correction values). Default parameters are
used if the long-press is done before any other module configuration.
CAUTION: The SDP installation method is designed for networks with
up to 15 modules. For larger sites, or better control over network paths,
please see the Smart-Vue Client software User Manual for setup instructions
or check with an experienced technician.
Follow these steps to install new modules automatically:
1. Press the button on the front of the module for 3 seconds.
2. If a receiver or repeater is within wireless range, the Smart-Vue module
automatically establishes the wireless connection, as shown in the
sequence of screen shots below:
Figure 73. Smart-Vue module automatic wireless setup sequence using
SDP feature