EasyManua.ls Logo

usi M9000 - Window

usi M9000
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
74
Verify Button
This button displays the Calibration Confirmation window. This function is intended
to be exercised at any time for an arbitrary verification of system calibration. This
feature can be used to test that the calibration for all input ranges of all analog
inputs has not changed by more than 0.25%. This is done by comparing
calibration factors calculated during this routine, with calibration factors that were
applied during the previous Internal Calibrate function.
Warning: The DME system will be taken off-line for the duration of this routine, typically 45
seconds.
Note: The Verify routine does not make any changes to the applied DME system calibration.
Figure 2-42 Calibration Confirmation Window
Yes Button
Selecting the Yes button activates the calibration verification routine.
Upon completion, the verification status report is displayed.

Table of Contents