962/964
SPECTROPHOTOMETER
3-1
3. Instrument Calibration
General Information 3-1
Positioning the Instrument
on the Reference 3-2
Calibration Procedure 3-3
General Information
Under normal circumstances, the instrument should be
calibrated at least once a day.
At the Main Menu, use the Tab Up $ or Tab Down @ key to
highlight Calibrate. Press the Enter key # to access the
Calibration Menu.
CALIBRATION
< Measure White Ref>
Status: Cal time up
S/N: ******
Ap Size: 8.0mm
The bottom portion of the calibration screen displays
information regarding the calibration status, cal plaque serial
number, and instrument aperture size. The status line displays
as either
Cal OK or Cal Required. Cal time up indicates that
calibration is required. Cal Ok indicates that no calibration is
required at this time. The serial number displayed on the
second line should match the serial number listed on your
calibration reference. The aperture size line displays the current
size. The line is also used to select the aperture size.
NOTE: Refer to Appendices for a detailed procedure on changing the
aperture setting and size.