Page #101 background imageLoading...
Page #101 background image
9 - 21
IM 4H3B1-01E
9. CHANGING INITIAL SETTINGS
9.6 How to Change the Temperature Unit (for USA)
For the TC or RTD type, the µR1800 will automatically append the temperature unit (˚C or ˚F).
The initial value is ‘˚C’, and common for all channels.
SETTING PROCEDURE:
Enter the SET UP Mode by turning ‘ON’ the power while pressing and holding the ENT-key
until the SET UP display appears.
Select ‘˚C’ or ‘˚F’ using the UP/DOWN-keys.
Press the ENT-key. The display ‘*TEMP SET*’ will appear.
This setting has been completed, but has not been stored yet. You can now adjust other settings in
the SET UP Mode, by using the ESC-key.
Before leaving the SET UP Mode, you have to store your new settings. Press the ESC-key to
return to the ‘SETUP=TEMP’ display and then select the ‘SETUP=END’ display.
Press the ENT-key. Select ‘END=STORE’ to keep your new settings or ‘END=ABORT’ and
press the ENT-key. After a few seconds, the Operation Mode will appear.
CAUTION
When you change this setting, all settings in the SET Mode will return to their initial
values. In the SET UP Mode, the RJC setting (9.2.3) and the input filter setting (9.2.4)
will return to their initial values. Other settings in the SET UP Mode will not be affected.
For a list, see 6.5
9.7 How to Initialize Settings
The settings which can be initialized concern the settings done in the SET Mode only.
This initialization does not affect any settings in the SET UP Mode.
For a list of initial settings of the SET Mode, see 6.5.
SETTING PROCEDURE:
Enter the SET UP Mode by turning ‘ON’ the power while pressing and holding the
ENT-key until the SET UP display appears.
Use the UP/DOWN-keys to select the display ‘SETUP=INIT’. Press the ENT-key.
Select ‘YES’ or ‘NO’ using the UP/DOWN-keys.
Press the ENT-key. The display ‘*INIT SET*’ will appear.
This setting has been completed, but, in case you selected ‘YES’, settings have not been
initialized yet. This will happen when you leave the SET UP Mode. You can now adjust
other settings in the SET UP Mode, by using the ESC-key.
When leaving the SET UP Mode, and you want to initialize your settings, press the
ESC-key. You will return to the ‘SETUP=INIT’ display and then select the
SETUP=END’ display.
Press the ENT-key. Select ‘END=STORE’ to initialize or ‘END=ABORT’ and press
the ENT-key. After a few seconds, the Operation Mode will appear.
9.6 How to Change the Temperature Unit (for USA)